model | The dimensions of the casing, including the installation and connection components | Internal dimensions | net weight | internal volume |
Width x Height x Depth [mm] | Width x Height x Depth [mm] | [kg] | [L] | |
MK56 | 720 x 1445 x 778 | 400 x 420 x 350 | 168 | 60 |
MK115 | 980 x 1725 x 865 | 600 x 480 x 400 | 260 | 115 |
MK240 | 1115 x 1715 x 925 | 735 x 700 x 443 | 340 | 228 |
MK720 | 1580 x 2005 x 1140 | 1200 x 1020 x 600 | 570 | 734 |
Temperature range: -40 ° C to 180 ° C
APT.line ™ Preheating chamber technology
Programmable condensation protection for samples
Heated observation window with LED internal lighting
Intuitive touch screen controller with time division and real-time functional programming
Internal data logger, measurement values can be read out when opened. Through USB interface format
Using capacitive humidity sensors and steam for humidity regulation and humidification
CFC refrigerant R-452A for MK and MKT series (refrigerant R-452A for MKF and MKFT series).
Fault diagnosis system using optical and sound alarms
Controller using interval programming
Access hole with silicone plug (model 53) 80mm, upper
Access hole with silicone plug (models 115, 240) 50mm, left
2 access holes with silicone plugs (model 720) 80mm, left and right
Independent adjustable temperature safety device level 2 (DIN 12880) with optical alarm
4 casters, 2 with brakes, starting from 115 liters
Computer interface (model 53): RS 422/(models 115, 240, 720): Ethernet
The socket on the right side operating component is 230 V, starting from 115 L
Display through LCD color monitor
Adjustable slope function function function
Integrated continuous automatic recorder
Real Time Clock
-
Door heating