KGF-1 Polymer Impurity Analysis System
1、 Basic configuration
1. UV detector 254nm
2. Constant flow plunger pump (flow rate range 0.1-10ml/min) pressure:<2MPa
3. Manual medium low pressure six way injection valve
4. High sensitivity chromatographic signal data processing system
2、 Instrument performance parameters
Sample elution flow<10ml/min (filler: dextran gel G-10, system test: blue dextran 2000)
Analysis flow rate: 0.1-5ml/min
Detector wavelength: 254nm, high sensitivity, high performance, dual beam.
Instrument stability time: Baseline stability within 10 minutes of startup.
Light source lifespan:>20000 hours
Injection quantitative tube: 200ul optional (standard 200ul)
Specialized glass polymer chromatography column (configured according to the 10th edition pharmacopoeia)
Minimum detection limit: 1 × 10-8g/ml; (Methanol solution of naphthalene)
Baseline noise: ≤ 1.0 × 10-5AU;? (≤10uv)
Baseline drift: ≤ 1 × 10-4AU/hr;? (≤200uv)
System repeatability testing accuracy
Repeated precision of 5 injections with 0.1mg/ml blue glucose: 1. Retention time difference ≥ 0.5%, 2. Peak height: ≥ 0.5%, 3. Peak area: ≥ 0.5%
3、 Instrument features
*The use of constant current and constant pressure medium and low pressure plunger pumps greatly improves the service life and testing accuracy.
*Specialized polymer impurity trace analyzer: It has the characteristics of high sensitivity, high performance, dual beam, long light source service life, long-term use without shutting down, and strong reliability.
*The equipped chromatographic acquisition and analysis software has functions such as data reanalysis and archiving, and the data results can be operated with Excl