Multi functional plasma dissolving machine (blood plasma dissolving machine)
Our company has independently developed a new multifunctional plasma melting machine to meet the needs of multiple hospital blood banks. The machine a
Product details

1Multi functional plasma dissolving machine (blood plasma dissolving machine) Main technical parameters and characteristics:

1. Temperature control 37 ℃± 1 ℃

2. Temperature control accuracy ± 0.5 ℃

3. Digital temperature display, microcomputer temperature control.

4. Heating power ≤1000W

5. Thawing time, ≤ 10 minutes

6. The maximum pulp volume is less than 8 bags (100-200ml)

7. There are 8 immovable stainless steel thawing blue pieces

8. Equipped with an automatic dual arm swing system that swings back and forth 60 times per minute or automatically adjusts the frequency.

IIMulti functional plasma dissolving machine (blood plasma dissolving machine)matters needing attention:

1. Normally set the water temperature, it can be turned on for a long time. When using it, turn on the swing switch, set the frequency, and it will automatically swing. After use, turn off the swing switch to save electricity and reduce mechanical wear.

2. Before use, be sure to add water to the water level line and avoid dry burning to protect the heating tube. As the startup time increases, the water in the water tank will gradually evaporate. Please check the water level at any time (add water and replace it). Ensure that the machine is operating normally.

3. Set temperature: Press the SET button to set or view the temperature set point. Press the SET key once to start flashing the character, indicating that the instrument has entered the set state. Press the △ key to increase the set value, and press the △ or ▽ key to quickly change the data. Press the SET key again to return the instrument to normal working state. Once the temperature setting is complete, the heating indicator light will turn on.

4. Pre setting: Press the SET button for 3 seconds to enter the inner layer parameter setting state of the instrument. *The parameter that appears and flashes is the advance amount that stops immediately. The advance amount parameter should be carefully adjusted to reduce temperature overshoot. When the instrument controls the heating output, it will cut off heating in advance. When the temperature drops below the advance amount, it will start heating again. The output (relay) will not work within the set value and advance amount range. This can reduce the number of relay actions and extend the service life of the relay. Example: If the set value is 50.0 ℃ and the advance amount is 0.5, the relay will release when the instrument control heats up to 49.5.0 ℃, and the relay will close again when the temperature drops to 50.0 ℃ -0.5 ℃=49.5 ℃. The larger the advance amount, the fewer times the relay will act, and if the advance amount is too large, it will reduce the control accuracy. After adjusting the advance parameter, press the SET button for 3 seconds to return the instrument to its working state. (Note: Pre setting is specifically set to prevent excessive temperature surge after heating stops, and generally does not require modification).

5. Error correction: When confirming that the value displayed on the instrument is not the correct measurement value, the displayed value can be corrected. Press the SET key for 3 seconds to enter the deep menu of the instrument panel. The * parameter that appears and flashes is E00, which is the advance quantity. Press the SET key again to display the flashing parameter, which is the error correction parameter. Use the △ or ▽ keys to modify this parameter. The range of error correction is correct, and the instrument is corrected to be incorrect. To reach -9.9 ℃ to+9.9 ℃, press the SET button again after the correction is complete to exit. The correction value of the instrument is 0.0 when it leaves the factory. When using it, it is necessary to prevent correcting the instrument that displays correctly to be incorrect.

Product related keywords: oscillator shaking table shaking bottle machine constant temperature box constant temperature oscillation machine