model |
3330 |
|
Measurement principles |
120 ° light scattering and membrane sampling |
|
Concentration limit |
Up to 3000 pieces/cm3 (3000000 pieces/L) |
|
mass concentration |
0.001—275000µg/m3 |
|
Particle size of particulate matter |
Detecting particle size range |
0.3—10µm |
Particle size resolution |
5% at 0.5 µ m (compliant with ISO21501-01/04) |
|
Particle size channel |
Up to 16 channels, user adjustable |
|
Time |
Greater than or equal to 1 second, user adjustable |
|
midnight |
<1 per minute |
|
internet traffic |
Sample |
1L/min ± 5% accuracy (compliant with ISO21501) |
Sheath gas |
1L/min |
|
environmental condition |
operate temperature |
0-45 degrees |
Operating humidity |
0-96%, no condensation |
|
storage temperature |
-20-60 degrees |
|
Aerosol medium |
air |
|
Data Storage |
5MB built-in storage (30000 samples) |
|
communication |
interface |
USB interface, Ethernet, and USB flash drive |
digital display |
5.7-inch color touch screen, drawing display |
|
Analog Output |
0-5V or 4-20mA, user selectable |
|
Alarm Output |
Visualization, buzzer, relay or switch quantity, user set alarm value |
|
Quality sampling |
Built in 37mm filter membrane |
|
light source |
Long life laser diode |
|
quantity of electricity |
ac adapter |
110 to 240VAC, 50/60Hz, 15W |
Size and quality |
volume |
13.5 x 21.6 x 22.4cm |
Weight (Instrument) |
1.6Kg |
|
Weight (battery) |
0.5Kg |
|
battery |
run time |
10 hours (1 piece) |
charging time |
4 hours (1 piece) |
|
safety appraisement |
CE |
EN61236-1:2006 |
CAN/CSA |
C22.2NO.61010-1 |
|
software |
Compatible with AIM software |
|
calibration |
NIST traceability, TSI recommends annual calibration |
|
Included Accessories |
Operation manual, power cord, battery, filter membrane, USB cable, packaging box, AIM software |
|
Options |
Backup battery, dual port external charger, diluter, external sound environment box |
Optical particle size spectrometer
Model:3330
The TSI3330 optical particle size spectrometer is simple and lightweight, capable of quickly and accurately measuring particle concentration and particle size distribution. Based on TSI's 40 year experience in aerosol instrument design, this product uses a 120 degree light scattering angle to collect scattered light intensity and a precise electronic processing system to obtain high-quality and high-precision data. At the same time, the strict calibration standards of TSI factory also ensure the accuracy of the instruments. TSI3330 can not only be used independently, but also placed in TSI's outdoor environment box for outdoor use.
application
Optical instruments are widely used in different fields due to their ease of operation, fast measurement speed, and durability.
Some of the main applications are as follows:
Filtering Test (ASHRAE52.2)
Indoor air quality
Workplace monitoring
Outdoor environmental monitoring
industrial measurement
Source emission monitoring and control
Features and advantages
Particle size resolution less than 5% at 0.5 microns
User adjustable particle size channel
Particle size range for detection: 0.3-10 microns, up to 16 channels
Detection concentration range: 0-3000/cm3
Color touch screen, intuitive user interface
Fully meets the requirements of ISO 21501-01/04
Inputting refractive index and density can simultaneously display particle number concentration and mass concentration
The collected sampling film can be used for weighing measurement and chemical analysis
The battery can provide a maximum of 20 hours of operation
Up to 30000 data can be stored
Record the temperature and pressure of each sample
Note: TSI3330 in the United States, TSI in the United StatesOptical particle size spectrometer 3330, TSI from the United States, acting as an agent for TSI particle size spectrometer 3330, and Zhongyi Environmental Science acting as an agent for TSI optical particle size spectrometer 3330, TSI dust analyzer, TSI anemometer, TSI laser particle counter, TSI flowmeter, TSI wind pressure gauge, TSI air hood, TSI particle image velocimeter.