Z4010 Zirconia Oxygen AnalyzerDetect the oxygen concentration in pure and inert mixed gases from the percentage level to below ppm. Its sturdy carrying case, fast response time, and switchable power supply voltage make it an ideal choice for on-site oxygen monitoring applications.
Use a large multi bit automatic shift display LCD to display the detected concentration and user adjustable alarm and analog output setting parameters. Programmable analog output provides an easy way to record results (if necessary).
Z4010 Zirconia Oxygen AnalyzerThere are two alarm channels, and users can set their status to high, low, off, and time delay. Each alarm is provided with a set of power-off conversion contacts. The alarm concentration can be set to any value within the instrument range. There are two types of analog outputs to choose from, 0-5V or 4-20mA, and there are several preset range options for either output. There is built-in sample gas flow regulation and monitoring, as well as a sampling pump used under low-pressure sample gas source conditions.
The standard sampling system consists of a simple needle valve and flow meter. The options include: sampling pump, fast flow loop system, and needle injection port. Hitech Instruments Co., Ltd. can also provide many other sample gas pretreatment accessories to meet the requirements of most sample gas sources.
Z4010 Zirconia Oxygen AnalyzerTechnical Parameter
display
Multi bitLCD -character height12.7mm
range
0.1ppm – 100%Oxygen, automatic shifting
Z4010 Zirconia Oxygen Analyzerdisplay resolution
Concentration in10%~100%The resolution is0.1%
Concentration in0.5%~9.99%The resolution is0.01%
Concentration in500ppm~4999ppmThe resolution is10ppm
Concentration in50ppm~499ppmThe resolution is1ppm
Concentration in0.00ppm~49.9ppmThe resolution is0.1ppm
accuracy
10%~100ppmReading of±2%Or better
99ppm~10ppm ±1ppm
0.1ppm~9.9ppm ±0.2ppm
stability
Better than readings every month2%perhaps0.5ppm-Take the bigger one.
Sample gas flow rate
The optimal operating flow is100~250ml/min
Response speed in a pure inert atmosphere
%Level:0.5sWithin to90%The amount of change;
follow1000ppmarrive10ppm:5sWithin to90%The amount of change;
100ppmfollowing:30sWithin to90%The amount of change;
When the oxygen sensor pool is stable/When adjusting at a low concentration level, the response to this concentration level is approximately3 sarrive4 s.
Sample gas pressure
The emptying pressure determines the pressure at the sensor. To achieve the accuracy specified in the technical data, the exhaust pressure should be atmospheric pressure.
Sample gas inlet pressure
maximum:6 bar
Sample gas temperature (at the analyzer)
highest100℃
Sampling system material
Including: nickel, copper, aluminum, stainless steel, platinum, zirconia, aluminaPTFENitrile rubber
Sample gas interface
Suitable for outer diameter0.25″(6mm)The sealed compression joint of the pipe.
Z4010 Zirconia Oxygen AnalyzerAnalog Output
Standard:0-5V
Optional:4-20mA
All are user programmable0-100%and0-5ppmbetween.
ambient temperature
The highest is45℃
power supply
On the front panel, you can choose:
110/120V、50/60Hz, maximum power consumption100VAperhaps
220/240V、50/60Hz, maximum power consumption100VA
Z4010 Zirconia Oxygen Analyzerinstall
IP67Fully waterproof suitcase