Scope of application:This product is suitable for microbiology laboratories in colleges, research institutes, and enterprises, and is an ideal device for precision fermentation experiments
Be prepared. Screening of culture medium formulas suitable for microbial fermentation cultivation, optimization of fermentation process parameters, and production process and strain selection
verification. The transparent glass design allows for easy observation of the materials inside the tank.
Technical Specifications:
|
Overview |
Technical Specifications |
|
Device Model |
HND-GJ-3L/5L/7L/10L/15L |
|
Diameter to height ratio |
1:2-2.5 |
|
Liquid loading coefficient |
65%-80% |
|
tank material |
Adopting high temperature and high pressure resistant borosilicate glass or a combination of stainless steel and borosilicate glass |
|
Sterilization method |
In situ or ex situ sterilization |
|
tank structure |
Transparent borosilicate glass body allows for intuitive observation of the state inside the tank. Various process interfaces, including:pHTheDO, temperature interface, sterile sampling port, feeding port, inoculation port, feeding port, foam electrode interface, pressure gauge port, air inlet, exhaust port, etc |
|
mixing system |
Adopting upper direct mechanical stirring or bottom magnetic stirring, the kettle is sealed with mechanical seals, various forms of high-performance stirring blades, defoaming blades, DC or AC speed regulating motors, and the speed is infinitely adjustable |
|
pressure control |
Manual control, local indication of pressure gauge, measurement and control range:0-0.4Mpa |
|
flow control |
Manual control, glass rotor flowmeter display |
|
temperature control |
Jacket, inner tube circulating water bath control or external heating device, with a control range of cooling water temperature+5℃~65℃±0.2℃ |
|
PHcontrol |
Swiss imported Mettler electrode (or Hamilton) online detection, peristaltic pump automatically adds acid or alkali for precise controlpHValue. Measurement range:0.00~14.00±0.01pHControl range:2.00~12.00±0.05pH |
|
DOcontrol |
Swiss imported Mettler electrode (or Hamilton), online detection, can be associated with speed and other related controls. Measurement range:0-100%or0-200%Display accuracy1%or0.1% |
|
Supplement control |
PIDFully automatic peristaltic pump switch control, automatic flow addition and metering |
|
Foam control |
Automatic detection of defoaming electrode, automatic flow of defoaming agent by peristaltic pump |
|
control system |
HND-BIO6000The online process control system adopts a color LCD touch screen+PLCForm, capable of online detection and control of fermentation process related parameters. The specific functions include: sterilization interface, parameter calibration, parameter setting, data curve graph, data report, fermentation batch number setting, online animation, alarm setting, which can be connected to the upper computer and printer to achieve remote monitoring and data storage and printing |
|
Extended features |
Air flow control, tank pressure control, exhaust gas controlCO2TheO2Detection and analysis, multi-channel feeding, weighing |