BMT964BT Ozone Analyzer
OZONE ANAL YZER BMT964BT is a portable version of the standard BMT 964 (please refer to the data sheet of the standard BMT 964). This instrument is sl
Product details

Product Introduction

OZONE ANAL YZER BMT964BT is a portable version of the standard BMT 964 (please refer to the data sheet of the standard BMT 964). This instrument is slightly larger than the standard BMT964. As an option, BMT 964 BT can be ordered with a built-in blowing unit (three-way solenoid valve and air pump) for fully automatic blowing and zeroing analyzer.

OZONE ANAL YZER BMT964BTIt's standardBMT 964Portable version (please refer to the standard)BMT 964The data table). This instrument is higher than the standardBMT964Slightly larger. For safety reasons, it is recommended to use a wall mounted wide range power supply24VDC.

BMT 964 BTThe throttle valve is equipped with a flow meter on the right side (sample gas inlet) and a catalytic box (catalytic ozone disruptor) on the left side, which can remove ozone from the sample gas before discharging it into the environment. Sample gas passes through3x5mm1 / 8''x3 / 16''PTFEThe fittings of the pipe enter the throttle valve and then enter through a replaceable glass felt air filterBMT964BTLeaveBMT964BTWhen the sample gas passes through the catalytic column, the ozone inside is destroyed.

As an option,BMT 964 BTYou can order a built-in blowing unit (three-way solenoid valve and air pump) for fully automatic blowing and zeroing analyzer.

-Dual beam UV photometer

-Long life mercury lamp

-High precision, with an error not exceeding0.5

-Pressure and temperature compensation

-Scope from2to600g / Nm3

-Built in sample gas filter

-Optional built-in blowing unit

-Relay contact control automatic zeroing

-Error Alert: Insufficient Brightness, Sample Tank Contamination, Out of Range, Error Summary

-Set functions through panel buttons or computer window software

-The display unit is:g/ Nm3,%wt / wtperhapsppmv

-High and low limit alarm

-Timer setting automatically resets to zero

-Pressure readingBarPsiTorrperhapsMPaIn units

- 4-20mAand0-10VIsolated output

- RS-232Interface (bidirectional)

-Measurement of Ozone Generator

-Ozone measurement in the laboratory

-On site ozone measurement

-Ozone experiment

-waste water treatment

-Industrial ozone treatment

measuring principle

Dual beam UV photometer(254nm

MTBF

The instrument includes a UV lamp65,000hExcluding UV lamps120,000h

Ultraviolet Lamp

Low pressure mercury lamp, long-life design

display

16Backlit letters of one characterLCD display screen

concentration range

50,100,200,300,400,500,600g / Nm3, optional unitsg / Nm3,%wt / wtandppmv

Optional Range

2,5,10,20g / Nm3, optional unitsg / Nm3,%wt / wtandppmv

accuracy

The maximum sum of errors after zeroing is: the sum of the measured values0.4+Range of measurement0.1

Repeatability error

The measured values0.2

response time

0.03s(Analog output),0.3s(Display)

zero-drift

0.2/Oh my god(non-cumulativeSex)

Maximum inlet pressure

2.5 bar

Maximum export pressure

absolutepressure1.15Bar

ambient temperature

0-50°CmistakeCondensation)

advancegas port

used forPTFEpipe3x5mm1/8“ x3 / 16”),Optional4x6mmPTFE

The built-in sample gas filter is a standard configuration

referral traffic

standardThe value is0.1to1L/min

temperature compensation

standard

Pressure Compensation

Optional with built-in test tube pressure sensor unit:BarPsiTorrMPa

signal output

4-20mA(Isolation,activation0-10V(Isolation)

Concentration alarm

tall/Low concentration alarm

control in

Set to zero(24V18mAIsolation)

control output

Relay contacts,28V0.5AIsolationLow light, color cup pollution, high/Low alarm, purge control(Lamp LowCuvette DirtyHigh AlarmLow AlarmPurge Control

Error report handling

Error relay:30V1ASummary of instrument malfunction warnings and errors: bulb low warning, bulb low error, bulb off error, colorimetric dish dirty warning, colorimetric dish dirty error, over range, over pressure

serial interface

RS-232, bidirectional,2400-38400Baud

AUTO ZERO

Equipped with optional external or internal purge pump and solenoid valve (e.gBMTBlowing device)

power supply

100-240VAC15VAThrough wall mounted power supply24VDC

Size (width)xtallxDeep)

144x72x230mm, DIN43700

weight

2.1kg

You can chooseEquipped with OZONE ANALYZER BMT964BTBuilt in internal cleaning unit, consisting of a built-in three-way solenoid valve and an air pump with a particle filter, used for fully automatic cleaning and zeroing of instruments. This option ensures that the colorimetric cup is rinsed with a colorimetric dish (zero gas) when the auto zero function is activated.

For field applications, we recommend using a soft carrying bag that should have enough space to accommodate itBMT964BTAnd its attachments.

otherBMTProduct (for detailed information, please refer to the corresponding data sheet):

- BMT 964C(Cabinet) can achieve true automatic independent application in harsh service environments

- BMT 964 AQandBMT 964 AQ-ONLINEUsed for measuring ozone in deionized water

- BMT 964AQ / HFUsed for containingHFMeasuring Ozone in Deionized Water

- OZONE-IN-OFF-GASdevice

- BMT 932Ozone monitoring device, used for monitoring in ambient airTLVMonitoring(13and6Channel)

-ozone generatorBMT 802XandBMT 803Used for small-scale production of ozone from oxygen