DWY-100 Impact Test Low Temperature Tank
The DLYS-108A petroleum product kinematic viscosity tester (4-hole) is manufactured in accordance with the requirements specified in the industry stan
Product details

DLYS-108Atype Petroleum product kinematic viscosity tester

DLYS-108AOverview of the application of the petroleum product kinematic viscosity tester:

DLYS-108AThe manufacturing of the kinematic viscosity tester for petroleum products conforms to the industry standards of the People's Republic of ChinaSY/T5651The requirements specified in the Technical Conditions for Petroleum Product Kinematic Viscosity Testers are applicable to the standards of the People's Republic of ChinaGB/T265According to the provisions of the "Determination of Kinematic Viscosity of Petroleum Products and Algorithm for Dynamic Viscometers", the determination of liquid petroleum productsRefers to Newtonian liquidsThe kinematic viscosity under a constant temperature condition.

DLYS-108AProduct characteristics of a petroleum product kinematic viscosity tester

1Adopting an intelligent temperature controller, the temperature control is fast, the response is fast, the overshoot is small, and the temperature control accuracy reaches±0.1℃。

2Adopting a hard glass cylinder and an insulated jacket cylinderDouble cylinderGood insulation performance and clear observation of the sample.

3Adopting a desktop and all-in-one design, the instrument has good overall integrity and is easy to use.

4Using an electric stirring device, the temperature inside the bathtub is uniform.

DLYS-108ATechnical specifications of the kinematic viscosity tester for petroleum products:

1Working power supply:AC 220V±10%50Hz±5%.

2Heating power: two levels,1000Wand650W.

3Mixing motor: power6W; Speed1200r/min.

4Temperature measurement range: room temperature~100℃。

5Temperature control accuracy:±0.1℃。

6Constant temperature bath: capacity20LForm, with two layers of cylinders inside and outsideDouble-layer cylinder.

7Operating environment: ambient temperature~+35℃, relative humidity<85%.

8Temperature sensor: with a division mark ofPt100Industrial platinum resistors.

9Power consumption of the whole machine: not greater than1800W.

10Capillary viscometer: Suitable capillary or counter current capillary can be equipped according to user requirements.