GWF-D1 Particle Analyzer

Performance characteristics:

● Meet the requirements of multiple national standards for medical devices such as the Chinese Pharmacopoeia, the United States Pharmacopoeia, the Drug Packaging Material Standards, and the infusion equipment GB8368-2018.

High performance laser light source and optical energy compensation circuit ensure the testing accuracy of various colorless and colored clear samples.

The injection slit and pipeline are made of imported 316L and PTFE materials, which can directly detect weak acidic and alkaline corrosive special solutions such as organic solvents and oil matrices

Automatic lifting system, suitable for detecting irregular packaging inspection products

High performance processor with nearly 10000 counting channels, achieving high-precision data acquisition.

Fully automatic calibration improves calibration efficiency and ensures accurate and reliable test data.

It can be calibrated and calibrated according to standards such as the Chinese Pharmacopoeia, the United States Pharmacopoeia, and ISO21510, and can meet the requirements of drug import and export testing.

Equipped with permission management function to ensure data security

Adopting the Android operating system, with a user-friendly interactive experience and simplified operation.

Built in database with massive storage and data statistical analysis capabilities.

RS232 and USB interfaces allow for external computer storage of test results, facilitating data classification and retrieval.

Visual dust door design to prevent environmental contamination of samples.

Technical Specifications:

Channel settings: Simultaneously display 64 particle size channels, customizable with thousands of particle sizes and an accuracy of 0.1 µ m

Particle size range: 1-500 µ m

Counting range: 0-9999999 particles

Injection volume: 0.2-1000ml (accuracy 0.1ml)

Injection volume accuracy:<± 0.5%

Injection speed: 5-80ml/min

Counting accuracy:<specified value ± 5%

Channel resolution:>95%

Relative standard deviation: RSD<1.5% (standard particles ≥ 1000 particles/ml)

Ultimate detection concentration: 18000 particles/ml

Mixing speed: 0-1000 revolutions per minute

Working temperature: 10-40 ℃

Power supply: 100-240VAC; 50/60Hz;<80W

Data output: pin printer, RS232 interface, USB storage interface


GWF-D1 Particle Analyzer

Main Features:

Meet the requirements of national standards of China Pharmacopeia、America Pharmacopeia、Pharmaceutical package standard、infusion devices GB8368.

Full volume test, simulated clinical infusion process with objective and true data.

Original packing and completely closed sampling to ensure no pollution in the process.

It adopts new type of injection pump to ensure continuous sampling, high precision and fast speed.

The liquid level automatic recognition function can distinguish the end point test without setting the sampling volume.

The high performance laser light source and light energy compensation circuit, ensure the test precision of various samples of colorless, colored sample.

The high performance processor, nearly 10000 counting channels, realize high precision data acquisition.

The fully automatic calibration, built-in kinds of calibration curves, meet the requirements of different standards.

The multi-level rights management for data security.

Android operating system, built-in huge storage, data statistical analysis function.

You can access the laboratory operation platform, network management.


Main Technical Index:

Channel Set up:64 Channel, thousands of different particle sizes can be free set up, accuracy 0.1µm

Test Range:1~500µm

Count Range:0~9999999 particle

Sample Volume:1ml~full volume, accuracy 0.1ml

Accuracy of Sample Volume: <0.5%

Sample Speed: 5~80ml/min, arbitrary set up

Counting accuracy: < required value5%

Channel Resolution: >95%

Relative Standard Deviation: RSD<1.5%(standard particle1000 particle/ml)

The best detection concentration: 0~18000 particle/ml

Mixing Speed: 0~1000r/min, impeller mixer of adjustable

Work Temperature:10~40℃

Power: AC220V10% ; 50Hz; <80W

Data output: Built-in Pin mini-printer; RS232 interface can connect computer; USB saved interface.