Performance characteristics:
Meets the insoluble particle detection requirements of the Chinese Pharmacopoeia and the United States Pharmacopoeia, and can directly detect the size and quantity of insoluble particles in injection solutions, sterile powders, and infusion equipment
There is a dedicated testing program for fat emulsion injection, which meets the requirements of the Chinese Pharmacopoeia and the Drug Standards of the Ministry of Health of the People's Republic of China for fat emulsion injection. The emulsion injection is determined by the photoresist method, and the test results are directly given
High performance laser light source and light energy compensation circuit ensure the testing accuracy of various colorless and colored clear samples
The injection slit and pipeline are made of imported 316L and PTFE materials, which can directly detect weak acidic and alkaline corrosive special solutions such as organic solvents and oil matrices
High performance processor with nearly 10000 counting channels, achieving high-precision data acquisition
Fully automatic calibration to improve calibration efficiency and ensure the accuracy and reliability of test data
L can calibrate and calibrate instruments according to standards such as CHP, USP, ISO21510, etc., which can meet the requirements of drug import and export testing
The audit tracking function ensures the integrity of electronic data, allowing for easy access to device usage information and current detection data at any time
Multi level permission management ensures data security and meets the requirements for data audit and tracking in computerized systems
Adopting the Android operating system, with a user-friendly interactive experience and simplified operation
Built in database with massive storage and data statistical analysis capabilities
RS232 and USB interfaces, can be connected to an external computer to store test results, facilitating data classification and retrieval
Technical Specifications:
Channel settings: 64 particle size channels, customizable with thousands of particle sizes and an accuracy of 0.1 µ m
Particle size range: 1-500 µ m (sensor can be selected in segments)
Counting range: 0-9999999 particles
Injection volume: 0.2-1000ml (accuracy 0.1ml)
Injection volume accuracy:<± 0.5%
Injection speed: 5-80ml/min
Counting accuracy:<specified value ± 5%
Channel resolution:>95%
Relative standard deviation: RSD<1.5% (standard particles ≥ 1000 particles/ml)
Limit detection concentration: 18000 particles/ml
Mixing speed: 0-1000 revolutions per minute
Working temperature: 10-40 ℃
Power supply: 100-240VAC; 50/60Hz;<80W
Data output: pin printer, RS232 interface, USB storage interface