Mixed Karl Fischer Moisture Analyzer
The Karl Fischer moisture determination method has been recognized by many international standards, such as ISO, ASTM, DIN, BS, and JIS, as the most a
Product details
2. Electrolytic calibration measurement system: Automatically performs titration calibration without human error, with high accuracy.
3. Integrated measuring cell: The volumetric Coulombic method shares a titration cup to ensure sealing and improve convenience.
4. Electrode replaceable diaphragm: The diaphragm of the electrode can be disassembled for easy cleaning and maintenance.
5. Three measurement modes: volumetric method, Coulomb method, or mixed method can all be performed in the same measurement cell.
6. Standard USB port: A USB flash drive can be used to store measurement methods and test results as electronic records.
measurement range: 10μg~300mg H2O.
accuracy: measurement1mgStandard water sample, with a relative standard deviation less than0.3%, Resolution: 0.1μg.
容量分析法(Volumetric Method) :
measurement range: 100μg~500 mg H2O (according toKFTitration of test drug).
burette: capacity: 10mLRepeatability: ±0.005mL.
Mixed analysis method(Hybrid Method) :
measurement range: 10μg~500mg H2O.
accuracy: measurement10mgStandard water sample, with a relative standard deviation less than0.3%, Resolution: 0.1μg.
Electrolytic calibration method(Electrolysis) :
measurement range: Volumetric Karl Fischer reagent titration degree1~5mg/mL.
accuracy: The relative standard deviation is less than1%.
interface: RS-232C,SS-BUS,LAN.
power supply: AC100~240V, 50/60HZ, 80W.
size: 141(W)x296(D)x383(H)mm.
weight: about4.5kg.