Beijing Zhongyi Yusheng Technology Co., Ltd
Home>Products>PT-7900D-II type fully automatic blowing and capturing device
PT-7900D-II type fully automatic blowing and capturing device
Product Introduction: PT-7900D-II Blowing and Capturing Device is a new generation of fully automatic water and soil integrated blowing and capturing
Product details

1-200H91IH3X1.png

PT-7900D-Type fully automatic blowing and capturing device

Instrument Introduction

PT-7900D-Type II blowing and capturingIt is a new generation fully automatic water and soil integrated blowing and capturing device independently developed by Beijing Zhongyi Yusheng Technology Co., Ltd. Adopting a seven inch touch screen display control, imported injection pump sampling, electronic cold trap using imported semiconductor refrigeration capture, imported switching valve injection,MFCElectronic flow control, can be loaded at full load once72 sample bottles, compatible with all domestic and imported GC and GCMS.

functional configuration

1. Can automatically run at full load72 samples, no need for personnel on duty;

2. The functions of account login, system self-test, fault alarm and prompt, temperature overload protection, leakage protection, etc. have improved the safety factor of instrument use;

3. All gas paths are treated with inertness, using imported electric switching valvesMFCMass flow controllerhigh-precisionelectron pressuresensingThe device,approveControl the carrier gas pressure and purge flow rate to achieve stable retention time;

4. Parameter settings for detection methods, real-time cloud display of working status, running time, traffic flow, etc.

5. Program controlled sample area, injection valve, sample transfer tubeCold trapSix temperature channels in the desorption zone of the water trap and capture tube, with a temperature rise rate of up to1800℃/min

6. The seven inch touch screen operation interface design is beautiful, practical, easy to operate, and perfectly realizes the human-computer interaction process;

7. Can start synchronouslyGCAnd workstationsAlternatively, it is also possibleImplementation andCounter control between GCs;

8. A brand new software control system that automatically performs functions such as blowing, trapping, desorption, injection, and backwashing with just one click;

9. This device is equipped with automatic sample mixing functionUOptional Type Purging Tube5ml25ml

10. The automatic blowback function of the entire blowing and injection system can effectively avoid cross contamination of different samples;

11. A water trap is equipped in front of the capture trap to reduce the impact of water vaporThe impact of GCMS.

12. This instrument is equipped with dedicated interfaces for various imported instruments,Can connect all models both domestically and internationallyGC、GCMS

13. Added log, help, stopwatch, and Chinese English interface switching functions to provide more convenience.

Application scope

The detection concentration of volatile organic compounds in the environment can reachPpb level;mainVolatile organic compounds applied in water and soilVOCs analysis,It can also be used for the analysis of volatile compounds in food.instrumentconform toThe following standards:

1. HJ639-2012 Determination of volatile organic compounds in water quality by blow sweep capture/gas chromatography-mass spectrometry method

2. HJ686-2014 Determination of volatile organic compounds in water quality by blow sweep capture/gas chromatography-mass spectrometry method

3. HJ605-2011 Determination of Volatile Organic Compounds in Soil and Sediments by Sweep Capture/Gas Chromatography Mass Spectrometry, etc.


Online inquiry
  • Contacts
  • Company
  • Telephone
  • Email
  • WeChat
  • Verification Code
  • Message Content

Successful operation!

Successful operation!