VIP member
R&S RTO1044 4G bandwidth 4-channel oscilloscope
Introduction: The RSRTO1044 oscilloscope has excellent signal fidelity, extremely high acquisition rate, and the world's first real-time digital trigg
Product details
R&S ® The RTO1044 oscilloscope has excellent signal fidelity, extremely high acquisition rate, and the world's first real-time digital triggering system. It has a bandwidth of 4 GHz, a maximum sampling rate of 10G/S, and can output 4 channels simultaneously. These oscilloscopes have hardware accelerated measurement and analysis capabilities. R&S ® RTO1044 can quickly troubleshoot signal faults and has hardware accelerated analysis capabilities, with a high-precision digital triggering system, R&S ® RTO1044 operation method, simpler and faster, truly convincing measurement accuracy

R&S ® RTO1044 is widely used in the following fields:
Trigger and decoding of serial protocol
Support multiple standards
The collection rate is extremely high, allowing for quick troubleshooting
Quick and convenient configuration
Flexible protocol triggering
The data display is clear and concise
Intuitive navigation function
R&S ® Advantages of RTO1044
Quickly troubleshoot signal faults
One million waveforms per second: accurate troubleshooting without guessing
The collection rate is extremely high, and there are no restrictions on the functionality
Support historical information viewing function, fast fault analysis
Hardware acceleration analysis function
The measurement speed is extremely high, and it is no exception for complex analysis functions
FFT spectrum analysis: powerful and easy to operate
Template testing: Quick configuration, reliable results
Advanced analysis function, capable of analyzing up to three waveforms simultaneously per channel
High precision digital triggering system
The triggering jitter is extremely low, which can achieve high-precision measurement
Within the full bandwidth range, the triggering sensitivity is extremely high
For trigger signals, use adjustable digital filters
The operation method is simpler and faster
Using an intuitive menu structure for faster operation
The control components are color coded for easier identification
Signal icons support drag and drop functionality
The dialog box adopts a semi transparent design, and the complete measurement signal can always be observed
Convincing measurement accuracy
Extremely low inherent noise and high measurement accuracy
Using a single core analog-to-digital converter for a larger dynamic range
Full bandwidth measurement, even for input sensitivity range ≤ 10mV/div
Extremely small gain and deviation errors independent of temperature
The isolation between channels is extremely high, with no crosstalk phenomenon

R&S ® RTO1044 is widely used in the following fields:
Trigger and decoding of serial protocol
Support multiple standards
The collection rate is extremely high, allowing for quick troubleshooting
Quick and convenient configuration
Flexible protocol triggering
The data display is clear and concise
Intuitive navigation function
R&S ® Advantages of RTO1044
Quickly troubleshoot signal faults
One million waveforms per second: accurate troubleshooting without guessing
The collection rate is extremely high, and there are no restrictions on the functionality
Support historical information viewing function, fast fault analysis
Hardware acceleration analysis function
The measurement speed is extremely high, and it is no exception for complex analysis functions
FFT spectrum analysis: powerful and easy to operate
Template testing: Quick configuration, reliable results
Advanced analysis function, capable of analyzing up to three waveforms simultaneously per channel
High precision digital triggering system
The triggering jitter is extremely low, which can achieve high-precision measurement
Within the full bandwidth range, the triggering sensitivity is extremely high
For trigger signals, use adjustable digital filters
The operation method is simpler and faster
Using an intuitive menu structure for faster operation
The control components are color coded for easier identification
Signal icons support drag and drop functionality
The dialog box adopts a semi transparent design, and the complete measurement signal can always be observed
Convincing measurement accuracy
Extremely low inherent noise and high measurement accuracy
Using a single core analog-to-digital converter for a larger dynamic range
Full bandwidth measurement, even for input sensitivity range ≤ 10mV/div
Extremely small gain and deviation errors independent of temperature
The isolation between channels is extremely high, with no crosstalk phenomenon
| performance index | ||
| Vertical system | ||
| Channel |
|
2 |
|
|
4 | |
| Bandwidth (-3 dB) at |
|
600 MHz |
| 50 Ω | ||
|
|
1 GHz | |
|
|
2 GHz | |
|
|
4 GHz | |
| Rise time (calculated value) |
|
583 ps |
|
|
300 ps | |
|
|
175 ps | |
|
|
100 ps | |
| Input impedance | 50 Ω ±1.5 % | |
| 1 M Ω± 1% with 15 pF (measured) | ||
| Input sensitivity | Maximum bandwidth across the entire range | 50 Ω: 1 mV/div to 1 V/Div |
| 1M Ω: 1 mV/div to 10 V/Div | ||
| Effective Bit Number (ENOB) of Analog to Digital Converter | Full range sine, frequency<-3 dB bandwidth | >7 digits (measurement) |
| acquisition system | ||
| Maximum sampling rate (real-time) | 10 G samples per second per channel | |
|
|
10 Gsa/s per channel | |
| 20 Gsa/s using two channels | ||
| memory depth | Standard configuration, with each channel/1 channel in working state |
|
|
|
||
| Maximum Upgrade (R&S)®RTO-B102 option), |
|
|
| Each channel/1 channel is in working state |
|
|
| Maximum collection rate | Continuous collection and display, 10 G samples/s, 1 k samples | 1 million waveforms per second |
| Super segmentation mode | <200 ns blind spot time | |
| Data Extraction Mode | For any combination of extraction mode and waveform algorithm, each channel can support up to 3 waveforms at most | Sampling, peak detection, high resolution, effective value |
| waveform algorithm | None, envelope, average | |
| Interpolation mode | Linear, Sin(x)/x, sample-and-hold | |
| horizontal system | ||
| Time Base Range | 25 ps/div to 50 s/div | |
| Time base accuracy | After supply/calibration | 2.5 ppm |
| After supply/calibration (R&S®RTO-B4 option) |
0.02 ppm | |
| Channel offset correction | ± 100 ns (real-time offset correction, channel to channel triggering - such as state detection offset correction) | |
| TRIGGER | ||
| Trigger Type | Edge, interference, pulse width, under amplitude, window, timeout, interval, slope, data clock, mode, status, serial mode, I2C, SPI, Optional: UART/RS-232, LIN, CAN, FlexRay | |
| sensitivity | Definition of Trigger Lag Parameter | Automatic or manual setting range: 0.1 div to 5 div |
| Detectable minimum pulse interference signal h | 100 ps | |
|
|
50 ps | |
| coupling mode | For the selected channel, RF suppression can be selected within the cutoff frequency range of 100 kHz to 50% of the analog bandwidth | |
| Waveform arithmetic operation | ||
| Types of Algebraic Operations | Arithmetic operations, logical operations, comparison operations, frequency domain, digital filters | |
| Hardware accelerated mathematical calculation function | +, –, *, 1/x, |x|, Derive, log10, ln, log2Scaling, FIR, FFT amplitude | |
| Analysis and measurement functions | ||
| Hardware acceleration analysis function | Spectrum, histogram, template testing, cursor | |
| Hardware accelerated measurement function | Amplitude measurement, time measurement | |
| General data | ||
| size | Width x Height x Depth | 427 mm x 249 mm x 204 mm |
| (16.81 in x 9.8 in x 8.03 in) | ||
| weight |
|
9.6 kg (21.16 lb) |
| monitor | 10.4 "LC TFT color touch screen, | |
| 1024 × 768 pixels (XGA) | ||
| connect | 1Gbit/s LAN, 4xUSB 2.0, GPIB (optional), DVI (for external monitor), external trigger | |
|
|
||||||
| probe | bandwidth | Attenuation factor | Input impedance | Input capacitance | dynamic range | other |
| passive probe | ||||||
|
|
500 MHz | 10:01 | 10 MΩ | ~10 pF | 400 V | |
| (Vrms) | ||||||
|
|
500 MHz | 10:01 | 10 MΩ | ~10 pF | 400 V | |
| (Vrms) | ||||||
| high voltage probe | ||||||
|
|
400 MHz | 100:01:00 | 50 MΩ | 7.5 pF | 1000 V | |
|
|
1000:1 | (Vrms) | ||||
| High voltage differential probe | ||||||
|
|
100 MHz | 100:1/1000:1 | 8 MΩ | 3.5 pF | ±140 V / ±1400 V | |
| Active Probe | ||||||
|
|
1.0 GHz | 10:01 | 1 MΩ | 0.8 pF | ±8 V | |
|
|
1.0 GHz |
|
||||
|
|
1.5 GHz | |||||
|
|
3.0 GHz | |||||
| differential probe | ||||||
|
|
1.5 GHz | 10:01 | 1 MΩ | 0.6 pF | ±5 V | Support high-precision DC voltmeter function; Front end micro buttons that support preset functions |
|
|
3.0 GHz | |||||
| Current probes | ||||||
| Probes | Bandwidth | Max. current | Rise time | Sensitivity error | Max. circuit voltage | Extras |
| (RMS / peak) | ||||||
| Current probes | ||||||
|
|
10 MHz | 150 A / ±300 A | 35 ns | ±1 % up to 150 A (RMS) | 600 V (CAT II) | external power supply necessary, |
| 300 V (CAT III) | e.g. R&S®RT-ZA13 |
|||||
|
|
100 MHz | 30 A / ±50 A | 3.5 ns | ±1 % up to 30 A (RMS) | 300 V (CAT I) | external power supply necessary, |
| e.g. R&S®RT-ZA13 |
||||||
Online inquiry