S-SCPH online pH sensor
● The online pH meter adopts high-precision AD conversion and single-chip microcomputer processing technology ● Multiple functions such as pH measurem
Product details
1. Product Introduction
1.1 Product Overview
S-SCPH is one of the intelligent online chemical analyzers developed by our company. It has strong environmental adaptability, clear display, simple operation, and excellent testing performance, making it highly cost-effective. It can accurately measure the pH value and temperature of solutions. It can be widely used for continuous monitoring of pH values in solutions such as thermal power, chemical fertilizers, metallurgy, environmental protection, pharmaceuticals, biochemistry, food, and tap water. Paired with ORP redox electrodes, it is also a precise ORP instrument.1.2 Functional Features
High intelligence: The online pH meter adopts high-precision AD conversion and single-chip microcomputer processing technology, which can complete various functions such as pH measurement, temperature measurement, and automatic temperature compensation.Time display: The Chinese online pH meter has a built-in clock chip that can display the current time.
Data storage: The built-in storage chip of the instrument can store historical data from the past three days, making it convenient for users to search.
N Rinse relay: The instrument is equipped with a rinse relay, which can be installed with an electromagnetic valve to regularly clean the pH electrode.
Conversion to 25 ℃: A 25 ℃ reference temperature conversion was performed for pure water and ammonia added ultrapure water, achieving the display of pH at 25 ℃
The n-value is particularly suitable for measuring various water qualities in power plants.
Strong anti-interference ability: using the latest devices with impedance up to 10-12 Ω; The current output adopts photoelectric coupling isolation technology, which has strong anti-interference ability and achieves remote transmission. Has good electromagnetic compatibility.
Simultaneous display of multiple parameters: Display time, pH or mV value, temperature, and output current value simultaneously on the LCD screen. Adopting a high brightness 128 * 64 LCD display module, it is eye-catching and has a long visible distance.
RS485 communication interface: MODBUS RTU communication protocol can be easily connected to a computer for monitoring and communication.
Industrial control watchdog: Ensure that the instrument does not crash.
1.3 Technical indicators
| Technical Specifications | technical indicators |
| execution standard | JB/T 6203-92 "Industrial pH Meter" |
| PH measurement range | 0-14.00pH maximum -2-16PH); Division value: 0.01pH |
| PH measurement accuracy | ±0.02pH |
| ORP measurement range | -1999~+1999mV; division value: 1mV |
| ORP measurement accuracy | ±1mV |
| Temperature measurement range | 0-99.9 ℃ (maximum -10~120 ℃); Division value: 0.1 ℃ |
| Temperature measurement accuracy | ±0.5℃ |
| Automatic temperature compensation range | 0-99.9 ℃, based on 25 ℃ |
| Repetitive error of electronic unit | ±0.02pH |
| stability | ±0.02pH/24h |
| input impedance | ≥1012Ω |
| Current isolation output | 4~20mA |
| Output current error | ≤±1%FS; |
| High and low alarm relay | AC220V、3A; |
| Data storage | 72 points, three-day data. |
| RS485 communication interface | MODBUS RTU protocol |
| power supply | AC220V ± 22V, 50 ± 1Hz; optional DC24V ± 2.4V |
| Protection level | IP54 |
| Overall dimensions | 96 (length) x 96 (width) x 110 (depth) mm |
| cutout size | 92×92 mm |
| Instrument weight | Transmitter: 0.8kg |
| working conditions | Environmental temperature: 0-60 ℃ Relative humidity:<85% |
| Electrode wire length | Standard 5-meter configuration |
| S- | Company Code Name | ||||
| SCPH- | PH/ORP controller | ||||
| N01- | 485 communication | ||||
| B- | Cabinet mounted square digital display shell | ||||
| 201 | Conventional composite electrode | ||||
2. Maintenance of electrodes
The quality of pH meter usage largely depends on the maintenance of the electrode. The electrodes should be cleaned regularly to ensure that they are not contaminated or blocked; Calibrate the electrodes periodically; During the water outage, it is necessary to ensure that the electrodes are immersed in the tested liquid, otherwise their lifespan will be shortened; The cable connectors must be kept clean and free from moisture or water ingress. It is indeed a problem with the instrument, please do not repair it yourself, please contact us.1. Activation: If the electrode is stored in a dry environment, it must be soaked for 24 hours before use to activate it, otherwise calibration and measurement will result in significant errors.
2. Cleaning: When the electrode is found to be contaminated or blocked, affecting the measurement accuracy, a soft brush can be used to gently brush the electrode head, and then clean it with clean water. The cleaning methods for various types of pollution are as follows:
| Inorganic metal oxides | Cleaning with less than 1mol/dilute acid |
| Organic oils and fats | Dilute detergent (weakly acidic) cleaning |
| Resin polymer material | Alcohol cleaning |
| Protein hemocyte sediment | Acid enzyme solution cleaning |
| Pigment substances | Dilute bleach cleaning |
3. Regeneration: When the electrode response is found to be slow and almost sluggish, soak it in a mixture of 10% HNO3 and NH4F (50g/l) for regeneration. Usually within 10 seconds.
At the factory, the electrode has a protective cap containing a sponge soaked in KCl solution to maintain the moisture of the sensitive membrane. When the electrode is not in use, rinse it with water and insert it back into the protective cap containing 3mol/L KCl solution, or soak the electrode in KCl solution.
5. 485 communication protocol
Note: This feature is optional and only available on devices with 485 functionality.
5.2 Definition of Data Frame Format
Adopting Modbus RTU communication protocol, the format is as follows:
The time for the initial structure to be ≥ 4 bytes
Address code=1 byte
Function code=1 byte
Data area=N bytes
Error check=16 bit CRC code
Time to end structure ≥ 4 bytes
Address code: It is the address of the transmitter and is unique in the communication network (factory default 0x01).
Function code: The instruction function indication issued by the host.
Data area: The data area contains specific communication data, please note that the high byte of 16 bits data comes first!
CRC code: a two byte checksum.
Host query frame structure:
Slave response frame structure:
8.3 register address
8.4 Example and Explanation of Communication Protocol
Example: Read device address0x01ofPHValue and temperature value
Inquiry frame
Response frame
Calculation of pH and temperature values:
PH value: 02BC (hexadecimal)=700 (decimal)=>PH value=7.00
Temperature value: 00FA (hexadecimal)=250 (decimal)=>Temperature value=25.0 ℃
9.Precautions and Maintenance
Secondary watches generally do not require daily maintenance. In case of obvious malfunctions, please do not open them for self repair and contact us as soon as possible!
After starting the power supply, the instrument should display. If there is no display or the display is abnormal, the power should be turned off immediately, and the power supply and fuse should be checked for normal operation.
The cable connector must be kept clean and free from moisture or water, otherwise the measurement will be inaccurate.
The electrodes should be cleaned regularly to ensure they are not contaminated.
Calibrate the electrodes periodically.
During the water outage, it is important to ensure that the electrodes are immersed in the tested liquid, otherwise their lifespan will be shortened. The quality of pH meter usage largely depends on the maintenance of the electrode
Note: This feature is optional and only available on devices with 485 functionality.
5.1 Basic Communication Parameters
| code | 8-bit binary |
| data bit | 8 digits |
| parity bit | none |
| stop bit | 1 person |
| Error checking | CRC (Redundant Cyclic Code) |
| Baud rate | Can be set, factory default is 4800bit/s |
Adopting Modbus RTU communication protocol, the format is as follows:
The time for the initial structure to be ≥ 4 bytes
Address code=1 byte
Function code=1 byte
Data area=N bytes
Error check=16 bit CRC code
Time to end structure ≥ 4 bytes
Address code: It is the address of the transmitter and is unique in the communication network (factory default 0x01).
Function code: The instruction function indication issued by the host.
Data area: The data area contains specific communication data, please note that the high byte of 16 bits data comes first!
CRC code: a two byte checksum.
Host query frame structure:
| address code | function code | Register starting address | register length | Low byte checksum | High byte checksum |
| 1 byte | 1 byte | 2 bytes | 2 bytes | 1 byte | 1 byte |
| address code | function code | effective byte count | Data Zone 1 | Data Zone 2 | Data Zone N | Low byte checksum | High byte checksum |
| 1 byte | 1 byte | 1 byte | 2 bytes | 2 bytes | 2 bytes | 1 byte | 1 byte |
| register address | PLC or configuration address | content | operation | Definition Description |
| 0000 H | 40001 (decimal) | PH value | read-only | PH real-time value (expanded by 100 times) |
| 0001 H | 40002 (decimal) | temperature value | read-only | Actual temperature value (increased by 10 times) |
|
Inquiry frame
Response frame
|
address code |
function code |
Return the number of valid bytes |
PH value |
temperature value |
Low byte checksum | High byte checksum |
|
0x01 |
0x03 |
0x02 |
0x02 0xBC |
0x00 0xFA |
0xBA |
0x2C |
PH value: 02BC (hexadecimal)=700 (decimal)=>PH value=7.00
Temperature value: 00FA (hexadecimal)=250 (decimal)=>Temperature value=25.0 ℃
9.Precautions and Maintenance
Secondary watches generally do not require daily maintenance. In case of obvious malfunctions, please do not open them for self repair and contact us as soon as possible!
After starting the power supply, the instrument should display. If there is no display or the display is abnormal, the power should be turned off immediately, and the power supply and fuse should be checked for normal operation.
The cable connector must be kept clean and free from moisture or water, otherwise the measurement will be inaccurate.
The electrodes should be cleaned regularly to ensure they are not contaminated.
Calibrate the electrodes periodically.
During the water outage, it is important to ensure that the electrodes are immersed in the tested liquid, otherwise their lifespan will be shortened. The quality of pH meter usage largely depends on the maintenance of the electrode
Interface display instructions:
First line: Display as current time;
Second line: displayed as the value for measuring PH or ORP;
The third line displays the current measured temperature value and the output 4-20mA current value;
Function button description:
Press the "MENU" key to enter the menu interface. Before entering, please enter the correct password. The default password is 0000.
Press the "ESC" key to return to the measurement mode
Press the "▲" key to scroll up to view parameter items or increase data.
Press the "▼" key to scroll down to view parameter items or reduce data.
Press the 'ENT' key to store the modified parameter values.
First line: Display as current time;
Second line: displayed as the value for measuring PH or ORP;
The third line displays the current measured temperature value and the output 4-20mA current value;
Function button description:
Press the "MENU" key to enter the menu interface. Before entering, please enter the correct password. The default password is 0000.
Press the "ESC" key to return to the measurement mode
Press the "▲" key to scroll up to view parameter items or increase data.
Press the "▼" key to scroll down to view parameter items or reduce data.
Press the 'ENT' key to store the modified parameter values.
3.1 Menu Structure
By pressing the MENU button on the transmitter of the online pH meter and entering the correct password, you can enter the menu interface. The menu is a single level menu, and the functions and related parameters are shown in the table.| serial number | menu item | setting | parameter range | initial value |
|
1 |
High alarm setting |
Set high alarm relay Activate the pH value of the device |
-20.00~20.00PH |
14.00 PH |
|
2 |
Low alarm setting |
Set low alarm relay Activate the pH value of the device |
-20.00~20.00 PH |
0.00 PH |
|
3 |
4mA setting |
Set 4mA output Current corresponds to pH value |
-20.00~20.00 PH |
0.00 PH |
|
4 |
20mA setting |
Set 20mA output Current corresponds to pH value |
-20.00~20.00 PH |
14.00 PH |
|
5 |
Delay setting |
Set high and low alarm delay Hysteresis value |
0.00~1.00 PH |
0.00 PH |
| 6 | Time Settings | Set the current time | current time | |
| 7 | temperature correction | Correct temperature value | -99.9~+99.9 ℃ | 0.00 |
|
8 |
Potential value correction |
Enter password | 0000-9999 | 9999 |
| Coefficient correction | 0.900%~1.100% | 1% | ||
| offset | -99mV~ +99mV | 0mV | ||
|
9 |
filter coefficient |
Set the filter coefficient value (The larger the value, the more stable it is) |
0~5 |
0 |
| 10 | Two-point calibration | PH meter calibration | ||
| 11 | PH calibration | Correct PH offset | -20.00~20.00PH | 0.00 PH |
|
12 |
Manual temperature compensation |
Temperature of two composite electrodes compensation value |
0.0~+99.9 ℃ |
25.0℃ |
|
13 |
Password Settings |
Set the menu to enter password |
0000-9999 |
0000 |
|
14 |
Communication Settings |
Set instrument number | 0-255 | 001 |
| Baud rate | 2400/4800/9600 can be set | 9600 | ||
| 15 | Factory settings | Restore factory settings | Password: 9999 | |
|
16 |
Current correction |
Correction of 4-20mA electricity Stream output offset |
-9.99~+9.99mA |
0.00mA |
|
17 |
historical data |
Record the past three days Data (1 point/h) |
||
|
18 |
Rinse relay |
PH flushing relay Switches and Interval Time |
||
| 19 | Measurement Mode | Set the display pH or | PH value | |
| 20 | ORP value |