S-SCPH online pH sensor
● The online pH meter adopts high-precision AD conversion and single-chip microcomputer processing technology ● Multiple functions such as pH measurem
Product details

1. Product Introduction

1.1 Product Overview

S-SCPH is one of the intelligent online chemical analyzers developed by our company. It has strong environmental adaptability, clear display, simple operation, and excellent testing performance, making it highly cost-effective. It can accurately measure the pH value and temperature of solutions. It can be widely used for continuous monitoring of pH values in solutions such as thermal power, chemical fertilizers, metallurgy, environmental protection, pharmaceuticals, biochemistry, food, and tap water. Paired with ORP redox electrodes, it is also a precise ORP instrument.

1.2 Functional Features

High intelligence: The online pH meter adopts high-precision AD conversion and single-chip microcomputer processing technology, which can complete various functions such as pH measurement, temperature measurement, and automatic temperature compensation.
Time display: The Chinese online pH meter has a built-in clock chip that can display the current time.
Data storage: The built-in storage chip of the instrument can store historical data from the past three days, making it convenient for users to search.
N Rinse relay: The instrument is equipped with a rinse relay, which can be installed with an electromagnetic valve to regularly clean the pH electrode.
Conversion to 25 ℃: A 25 ℃ reference temperature conversion was performed for pure water and ammonia added ultrapure water, achieving the display of pH at 25 ℃
The n-value is particularly suitable for measuring various water qualities in power plants.
Strong anti-interference ability: using the latest devices with impedance up to 10-12 Ω; The current output adopts photoelectric coupling isolation technology, which has strong anti-interference ability and achieves remote transmission. Has good electromagnetic compatibility.
Simultaneous display of multiple parameters: Display time, pH or mV value, temperature, and output current value simultaneously on the LCD screen. Adopting a high brightness 128 * 64 LCD display module, it is eye-catching and has a long visible distance.
RS485 communication interface: MODBUS RTU communication protocol can be easily connected to a computer for monitoring and communication.
Industrial control watchdog: Ensure that the instrument does not crash.

1.3 Technical indicators

Technical Specifications technical indicators
execution standard JB/T 6203-92 "Industrial pH Meter"
PH measurement range 0-14.00pH maximum -2-16PH); Division value: 0.01pH
PH measurement accuracy ±0.02pH
ORP measurement range -1999~+1999mV; division value: 1mV
ORP measurement accuracy ±1mV
Temperature measurement range 0-99.9 ℃ (maximum -10~120 ℃); Division value: 0.1 ℃
Temperature measurement accuracy ±0.5℃
Automatic temperature compensation range 0-99.9 ℃, based on 25 ℃
Repetitive error of electronic unit ±0.02pH
stability ±0.02pH/24h
input impedance ≥1012Ω
Current isolation output 4~20mA


Output current error ≤±1%FS;
High and low alarm relay AC220V、3A;
Data storage 72 points, three-day data.
RS485 communication interface MODBUS RTU protocol
power supply AC220V ± 22V, 50 ± 1Hz; optional DC24V ± 2.4V
Protection level IP54
Overall dimensions 96 (length) x 96 (width) x 110 (depth) mm
cutout size 92×92 mm
Instrument weight Transmitter: 0.8kg
working conditions Environmental temperature: 0-60 ℃ Relative humidity:<85%
Electrode wire length Standard 5-meter configuration
1.4 Product Selection
S- Company Code Name
SCPH- PH/ORP controller
N01- 485 communication
B- Cabinet mounted square digital display shell
201 Conventional composite electrode

2. Maintenance of electrodes

The quality of pH meter usage largely depends on the maintenance of the electrode. The electrodes should be cleaned regularly to ensure that they are not contaminated or blocked; Calibrate the electrodes periodically; During the water outage, it is necessary to ensure that the electrodes are immersed in the tested liquid, otherwise their lifespan will be shortened; The cable connectors must be kept clean and free from moisture or water ingress. It is indeed a problem with the instrument, please do not repair it yourself, please contact us.
1. Activation: If the electrode is stored in a dry environment, it must be soaked for 24 hours before use to activate it, otherwise calibration and measurement will result in significant errors.
2. Cleaning: When the electrode is found to be contaminated or blocked, affecting the measurement accuracy, a soft brush can be used to gently brush the electrode head, and then clean it with clean water. The cleaning methods for various types of pollution are as follows:
Inorganic metal oxides Cleaning with less than 1mol/dilute acid
Organic oils and fats Dilute detergent (weakly acidic) cleaning
Resin polymer material Alcohol cleaning
Protein hemocyte sediment Acid enzyme solution cleaning
Pigment substances Dilute bleach cleaning

3. Regeneration: When the electrode response is found to be slow and almost sluggish, soak it in a mixture of 10% HNO3 and NH4F (50g/l) for regeneration. Usually within 10 seconds.
At the factory, the electrode has a protective cap containing a sponge soaked in KCl solution to maintain the moisture of the sensitive membrane. When the electrode is not in use, rinse it with water and insert it back into the protective cap containing 3mol/L KCl solution, or soak the electrode in KCl solution.

5. 485 communication protocol
Note: This feature is optional and only available on devices with 485 functionality.

5.1 Basic Communication Parameters

code 8-bit binary
data bit 8 digits
parity bit none
stop bit 1 person
Error checking CRC (Redundant Cyclic Code)
Baud rate Can be set, factory default is 4800bit/s
5.2 Definition of Data Frame Format
Adopting Modbus RTU communication protocol, the format is as follows:
The time for the initial structure to be ≥ 4 bytes
Address code=1 byte
Function code=1 byte
Data area=N bytes
Error check=16 bit CRC code
Time to end structure ≥ 4 bytes
Address code: It is the address of the transmitter and is unique in the communication network (factory default 0x01).
Function code: The instruction function indication issued by the host.
Data area: The data area contains specific communication data, please note that the high byte of 16 bits data comes first!
CRC code: a two byte checksum.
Host query frame structure:

address code function code Register starting address register length Low byte checksum High byte checksum
1 byte 1 byte 2 bytes 2 bytes 1 byte 1 byte
Slave response frame structure:
address code function code effective byte count Data Zone 1 Data Zone 2 Data Zone N Low byte checksum High byte checksum
1 byte 1 byte 1 byte 2 bytes 2 bytes 2 bytes 1 byte 1 byte
8.3 register address
register address PLC or configuration address content operation Definition Description
0000 H 40001 (decimal) PH value read-only PH real-time value (expanded by 100 times)
0001 H 40002 (decimal) temperature value read-only Actual temperature value (increased by 10 times)
8.4 Example and Explanation of Communication Protocol


address code

function code

starting address

data length

Low byte checksum

High byte checksum

0x01

0x03

0x00 0x00

0x00 0x02

0xC4

0x0B
Example: Read device address0x01ofPHValue and temperature value
Inquiry frame




Response frame


address code


function code
Return the number of valid bytes
PH value


temperature value
Low byte checksum High byte checksum

0x01

0x03

0x02

0x02 0xBC

0x00 0xFA

0xBA

0x2C
Calculation of pH and temperature values:
PH value: 02BC (hexadecimal)=700 (decimal)=>PH value=7.00
Temperature value: 00FA (hexadecimal)=250 (decimal)=>Temperature value=25.0 ℃
9.Precautions and Maintenance
Secondary watches generally do not require daily maintenance. In case of obvious malfunctions, please do not open them for self repair and contact us as soon as possible!
After starting the power supply, the instrument should display. If there is no display or the display is abnormal, the power should be turned off immediately, and the power supply and fuse should be checked for normal operation.
The cable connector must be kept clean and free from moisture or water, otherwise the measurement will be inaccurate.
The electrodes should be cleaned regularly to ensure they are not contaminated.
Calibrate the electrodes periodically.
During the water outage, it is important to ensure that the electrodes are immersed in the tested liquid, otherwise their lifespan will be shortened. The quality of pH meter usage largely depends on the maintenance of the electrode

Interface display instructions:
First line: Display as current time;
Second line: displayed as the value for measuring PH or ORP;
The third line displays the current measured temperature value and the output 4-20mA current value;
Function button description:
Press the "MENU" key to enter the menu interface. Before entering, please enter the correct password. The default password is 0000.
Press the "ESC" key to return to the measurement mode
Press the "▲" key to scroll up to view parameter items or increase data.
Press the "▼" key to scroll down to view parameter items or reduce data.
Press the 'ENT' key to store the modified parameter values.

3.1 Menu Structure

By pressing the MENU button on the transmitter of the online pH meter and entering the correct password, you can enter the menu interface. The menu is a single level menu, and the functions and related parameters are shown in the table.

serial number menu item setting parameter range initial value

1

High alarm setting
Set high alarm relay
Activate the pH value of the device

-20.00~20.00PH

14.00 PH

2

Low alarm setting
Set low alarm relay
Activate the pH value of the device

-20.00~20.00 PH

0.00 PH

3

4mA setting
Set 4mA output
Current corresponds to pH value

-20.00~20.00 PH

0.00 PH

4

20mA setting
Set 20mA output
Current corresponds to pH value

-20.00~20.00 PH

14.00 PH

5

Delay setting
Set high and low alarm delay
Hysteresis value

0.00~1.00 PH

0.00 PH
6 Time Settings Set the current time current time
7 temperature correction Correct temperature value -99.9~+99.9 ℃ 0.00


8


Potential value correction
Enter password 0000-9999 9999
Coefficient correction 0.900%~1.100% 1%
offset -99mV~ +99mV 0mV

9

filter coefficient
Set the filter coefficient value
(The larger the value, the more stable it is)

0~5

0
10 Two-point calibration PH meter calibration
11 PH calibration Correct PH offset -20.00~20.00PH 0.00 PH

12

Manual temperature compensation
Temperature of two composite electrodes
compensation value

0.0~+99.9 ℃

25.0℃

13

Password Settings
Set the menu to enter
password

0000-9999

0000

14

Communication Settings
Set instrument number 0-255 001
Baud rate 2400/4800/9600 can be set 9600
15 Factory settings Restore factory settings Password: 9999

16

Current correction
Correction of 4-20mA electricity
Stream output offset

-9.99~+9.99mA

0.00mA

17

historical data
Record the past three days
Data (1 point/h)

18

Rinse relay
PH flushing relay
Switches and Interval Time
19 Measurement Mode Set the display pH or PH value
20 ORP value
After the installation of the online pH meter is completed, the operator can calibrate the pH meter through the 10th or 11th menu and it can work normally:It is recommended to use the default factory settings for other menus, and it is best not to make corrections on your own. Please adjust the settings according to the actual situation under the guidance of professional personnel.