This instrument is designed in accordance with the standards of the People's Republic of ChinaGB 7600Determination of Moisture Content in Running Transformers (Coulombic Method), Standard of the People's Republic of ChinaGB/T11133Determination of Water Content in Liquid Petroleum Products (Karl Fischer Method), Industry Standard of the People's Republic of ChinaSH/T0246The method for determining the moisture content in light petroleum products (electric method) is designed and manufactured in accordance with the requirements specified in the above standards, and is applicable to the determination of trace moisture content in liquid petroleum products using the methods specified in the above standards.
This instrument adopts microcomputer control technology, which has the characteristics of fast analysis speed, high accuracy, LCD screen display, automatic printing, and has functions such as instrument fault self diagnosis and menu selection; It is a fully functional, easy to operate, and fully automatic measuring and analyzing instrument.
1、 Main technical features
1This instrument adopts microcontroller control technology, which enables seamless sampling, testing, calculation, analysis, and result output, with a high degree of automation.
2This instrument has the characteristics of fast measurement speed, high accuracy, accurate and reliable data, and simple operation. The test results are automatically displayed and printed.
3This instrument has a desktop structure and adopts LCD display technology. Input parameters and output results are displayed on the LCD screen, which is intuitive, eye-catching, and easy to operate.
2、 Main technical parameters and indicators
1Measurement method:Electricity method;
2Display:LCD(240X128LCD Chinese character display,currentI(Yaxis),timeT(Xaxis)Draw the electrolysis rate curve;
3Electrolysis control: automatic electrolysis current control (maximum)400mA);
4Measurement range:1Microgram of water~100Milligrams of water;
5Electrolysis speed:2.0Milligrams of water per minute (maximum);
6Sensitivity threshold:0.1Micrograms of water;
7Printing: Wide line micro printer;
8Working power supply: AC220V±10%、50Hz;
9Power consumption of the whole machine:30W;
10Environmental temperature for use:5~40℃;
11Operating environment humidity:<80%;
12Net weight: approximately6.5Kilogram.