Shanghai Jingxi Instrument Manufacturing Co., Ltd
Home>Products>SYD-3536-I Cleveland Open Flash Point Tester
Product Groups
Firm Information
  • Transaction Level
    VIP member
  • Contact
  • Phone
    18916955498
  • Address
    No. 50, Lane 300, Shunda Road, Nanxiang Town, Jiading District, Shanghai
Contact Now
SYD-3536-I Cleveland Open Flash Point Tester
SYD-3536-I Cleveland Open Flash Point Tester
Product details

brand:Shanghai Jingxi Instrument Manufacturing Co., Ltd


1、 Purpose:

This instrument is suitable for use according to national standardsGB/T3536-83The method for determining the flash point and ignition point of petroleum products (Cleveland Open Cup method) determines that except for fuel oil, the open flash point is less than 79The flash point and ignition point of all petroleum products other than those with a temperature of ℃ are also applicable to the international standard ISO2592Determination of Flash Point and Ignition Point of Petroleum Products (Cleveland Open Cup Method). Compliant with SY/T5645Product standards.

2、 Applicable working conditions for the instrument

2.1 The instrument should be placed on a flat workbench with ambient temperature5-35 ℃.

2.2 The rated voltage of the instrument is 220V ± 10%, and the frequency is 50Hz ± 1Hz.

2.3 The studio power supply must have a good grounding device.

3、 Main technical parameters and indicators

Cleveland Oil Cup:

materialH62

inside diameter 63.5±0.5

Ignition device:

Ignition Source Butane gas, coal gas, liquefied gas, etc

Igniter aperture 0.8

Flame ball diameterφ4

Electric furnace:

Heating power:400W

Adjustable equivalent output voltage:0~220V

thermometer:

conform toGB/T514, Scale (-6~400) ℃, minimum scale 2 ℃

4、 Performance characteristics

4.1 Sealed electric furnace, with a long service life and rapid heating up;

4.2 Control the heating power through voltage adjustment, with a voltage setting range of 0-2220V;

4.3 Equipped with mains power compensation function (not affected by external voltage changes);

4.4 Semi automatic remote ignition, accurate ignition;

4.5 Instrument integrated gas pipe rack, which can use various gases such as coal gas, butane gas, liquefied gas, etc. for ignition;

4.6 Good repeatability of measurement results.

Online inquiry
  • Contacts
  • Company
  • Telephone
  • Email
  • WeChat
  • Verification Code
  • Message Content

Successful operation!

Successful operation!

Successful operation!