- product description
Gases that meet specific purposes, including single gases or mixed gases. There are 259 types of single gases, including 115 electronic gases, 63 organic gases, 35 inorganic gases, 29 halogenated carbon gases, and 17 isotopic gases.
Special gas analysisThe helium ion gas chromatograph is used for the analysis of trace impurities in specialty gases. The PDIHID helium ion gas chromatograph technology is employed to provide our customers with accurate results and meet the requirements of easy operation and maintenance. The user-friendly interface is used as a guide, allowing operators to apply it freely. The helium ion gas chromatograph can provide you with good protection at any time
1、 Special gas analysisAccording to the standard:
GB/T 28726-2012Gas analysis - Helium ionization gas chromatography method
2、 Special gas analysisPerformance indicators:
1Temperature control indicators:
Column box: 8 ℃ -400 ℃ at room temperature with an accuracy of ± 0.1 ℃
Sample injector: Accuracy ± 0.1 ℃ at room temperature between 8 ℃ and 400 ℃
Detector: Accuracy ± 0.1 ℃ at room temperature between 8 ℃ and 400 ℃
Temperature control quantity: eight channels
Program heating: 8th order
Heating rate: 0 ℃~40 ℃
Cooling rate: within 7 minutes (350 ℃ to 50 ℃)
2Helium ionization detector (PDD):
Discharge mode: Pulse discharge
Baseline noise: ≤ 0.1mv
Baseline drift: ≤ 0.5mv/30min
Detection limit: ≤ 1.0 × 10-10g/ml
3Carrier gas purifier:
Purified gases: He, Ar, Ne, Kr, Xe, Rn
Operating pressure: 1000Psi
Removing gases: H2, O2, N2, CO, CO2, CH4, H2O, NO, NH3, CF4, etc
Residual concentration: ≤ 10pppb
4Workstation performance:
Communication interface: RS232/USB interface
High precision: 24 bit high-precision A/D
Resolution: ± 1UV
Input voltage range: -5mv to+1v
Sampling frequency: 20 times/second
Dynamic range: 106
Integral sensitivity: 1 μ v • sec
Linearity:<± 0.1%. < span=''>
Reproducibility: 0.06%.
5Other parameters: size, weight, power supply:
Dimensions: Width 655mm x Height 500mm x Depth 480mm
Weight:~48kg
Power supply: 220V ± 22V, 50Hz, power ≥ 2kW
3、 Special gas analysis instrumentSystem configuration:
(1) GC-9560 gas chromatograph (2) helium ionization detector (PDHID)
(3) Central cutting system (4) multi column box system
(5) Helium Purifier (6) Standard Gas
(7) Chromatography Column (8) Oxygen Adsorption and Reduction System
(9) Special pressure reducing valve for carrier gas (10), no dead volume dedicated sampling valve
(11) GC-9560V4.0 Reverse Chromatography Workstation