VIP member
The pH online analyzer has complete functions, stable performance, and is compatible with various electrodes
The phg-2091Ax online analyzer is fully functional
Product details
Environmentally friendly sewage pH meter, equipped with 2 composite pH electrodes and 3 composite pH electrodes, with a 10 meter cable and a 2-meter installation bracket
PHG-2091AX Intelligent TypeIndustrial PH/ORP meter
Overview
pHG-2091AXThe industrial pH/ORP meter is an online pH (ORP) value measuring instrument for industrial use, widely used for online pH measurement of high-purity water and water production equipment in industries such as electronics, chemical, environmental protection, medicine and food. It is equipped with a pipeline connected composite electrode, which is easy to use, accurate in reading, and stable in numerical value.
Main technical indicators of the instrument
- Product Inspection Standard: JB/T6203
- Measurement range: 0.00~14.00PH, with a division value of 0.01;
- Potential value: -1999~+1999mv, division value 0.1mv;
- Temperature value: 0~99.9℃, with a division value of 0.1 ℃;
- Accuracy: ± 0.05(PH);
- Temperature compensation: automatic temperature compensation, compensation range 0~99.9 ℃, compensation reference 25 ℃, temperature compensation element Pt1000;
- Signal output: 4-20mA with isolation, adjustable operating range, * high load 750 Ω;
- Display screen: Select a 56 × 38 (mm) large screen dot matrix backlit LCD display; Chinese menu style prompt operation;The instrument displays PHORP value, temperature value, current output current value, etc;
- Electrode configuration: PH/ORP composite electrode, 6 'pipe type threaded connection, lead length of 10m, * high pressure capacity of 0.6Mpa;
- PH/ORP value upper and lower limit alarm setting and output: This instrument has PH or ORP value upper and lower limit alarm settings, providing switch contact output when the set value is exceeded or exceeded, with a contact capacity of 220VAC/6A;
- RS485 communication output;
- Timed cleaning function: The instrument has timed and periodic light output signals, which are used to connect the cleaning electrode equipment;
- Instrument working conditions:
Environmental temperature: 0~50.0 ℃
relative humidity≤85%
- Electronic unit size:
Hole size: 91 × 91 (mm)
Environmental sewage detector PH meter
Chemical plant pH analyzer - resistant to strong acid-base corrosion
Online inquiry