WZ-1030E single-phase AC standard source
The WZ-1030E single-phase AC/DC standard source is a calibration source for LED five and a half digit display of AC/DC voltage and current. High outpu
Product details

The WZ-1030E single-phase AC/DC standard source is a calibration source for LED five and a half digit display of AC/DC voltage and current. High output power and strong load capacity; Equipped with an RS232 interface, it can be connected to a computer network. The system software design is reasonable and complies with the new national metrology regulations; Implement automatic processing of test data; The unique protection circuit makes the operation of the calibrator safer and more reliable.

This calibrator is mainly suitable for verifying and calibrating various current and voltage meter heads and corresponding precision multimeters below level 0.2. It can also be used as a high stability testing power supply, and can be used in conjunction with a high-grade standard meter to calibrate current and voltage meters above level 0.2.

product name

Single phase precision testing power supply, single-phase AC/DC standard source, single-phase programmable precision testing power supply, precision AC/DC testing power supply, single-phase AC/DC standard source.

Product Diagram

WZ-1030E单相交直流标准源-图解.jpg

Product characteristics

1. 5 1/2 digit LED digital display output, which can be displayed in two ways: actual value or percentage.

2. The output range of AC and DC currents is: 0-100uA~500uA~2mA~5mA~20mA~50mA~200mA~500mA~2A~5A~20A.

3. The output range of AC and DC voltage is: 0-200mV~1V~2V~5V~10V~20V~50V~100V~200V~500V~1000V.

4. There are three AC frequencies available for selection: 50Hz, 60Hz, and 400Hz (with crystal oscillator to ensure frequency stability).

5. Output overload can be automatically protected and manually reset.

6. There is a 10% margin for each voltage and current output range.

7. Clamp gauge measurement: Using our factory's calibration coil, it can measure 0-750A (some gauges can measure above 1000A); Conversion error ± 0.8%.

Product parameters

1. Stability: DC ≤ 0.01%/2 minutes of full scale

AC0.03%/2 minutes at full scale

When making precise measurements, the instrument needs to be preheated for two hours

2. Communication distortion:0.5%

3. DC ripple factor:0.05%

4. Output frequency and error range: 50Hz, 60Hz, 400Hz1%

5. The range and accuracy of output voltage and current (23 ℃± 2 ℃, output value greater than 10% range) are shown in the following table:

Output items

Output range

Rated Output

maximum output

accuracy

AC/DC

Voltage

0~200mV

10mA

30mA

DC: ± (0.03% of reading)

+0.02% of full scale

AC: ± (0.05% of reading)

+0.03% of full scale

Accuracy will not be assessed below ACV100 mV and ACI 2mA

0~2V

20mA

60mA

0~5V~10V~20V

40mA

100mA

0~200V~500V

20mA

60mA

0~1000V

10mA

30mA

AC/DC

electric current

0~100uA~500uA~2mA~5mA~20mA~50mA

DC:3V

AC:36V


0~200mA~500mA

DC:3V

AC:12V


0~2A~5A~20A

2V


Standard

Resistance

10. 24, 50, 100, 240, 500 (Ω) 1, 2.4, 5, 10, 24 (K Ω) plus x 1000, x 1 magnification levels (i.e. 10K Ω~24M Ω)

0.25W


±0.3%+20mΩ

6. Power consumption: AC power supply voltage 220V ± 10%, frequency 50Hz ± 1Hz; power consumption180VA

7. Working environment: The temperature of the working environment is between 5 ℃ and 35 ℃, and the relative humidity is not greater than 80%

8. Working hours: continuous

9. External dimensions: width x height x depth 480 x 135 x 420 (mm)

10. Weight: Approximately 16kg

Online inquiry
  • Contacts
  • Company
  • Telephone
  • Email
  • WeChat
  • Verification Code
  • Message Content

Successful operation!

Successful operation!

Successful operation!