VIP member
Product details
product image
1. Product IntroductionThe ZQ-YFT2002 desktop liquid pressure pump (hereinafter referred to as the pressure pump) is an important auxiliary equipment in pressure instrument detection work. It works in conjunction with a pressure standard to complete pressure instrument calibration and detection work. The high-pressure desktop liquid pressure pump material selects high-quality imported hydraulic pumps as pressure generating components, supplemented by well-designed fine-tuning and fast interfaces, to form a high-quality wide range pressure generating system with simple operation, reasonable structure, and good sealing performance. It can be widely used in industries such as metrology, scientific research, industrial and mining enterprises, and instrument manufacturing.
2. Main technical indicators
1) Pressure range
YFT2002A 0~40.0MPa
YFT2002B 0~60.0MPa
YFT2002C 0~100.0MPa
YFT2002Q 0~70MPa (medium is water)
2) Adjust sensitivity: min 10Pa
3) Dimensions: 400 × 360 × 160 (length × width × height)
4) Weight
YFT2002A 4kg
YFT2002B 5kg
YFT2002C 6kg
YFT2002Q 6kg
| serial number | name | function |
| 1 | Output 2 | Connect standard table or inspected table |
| 2 | pressure regulating valve | Lock when output pressure is unstable, ensuring stable pressure |
| 3 | Drain valve | Lock during pressurization and open during backpressure |
| 4 | lever | Drive the piston injection motion through operation |
| 5 | Oil storage chamber | Storage (usually using silicone oil) of pressure guiding liquid |
| 6 | Air release port | Open and release air during liquid injection |
| 7 | fine-tuning | Small scale increase or decrease in pressure |
| 8 | Output 1 | Connect standard table or inspected table |
| 9 | Pump body assembly | Combination installation body of various components |
| 10 | Lever lock buckle | Pull out or lock the fixed lever state |
| 11 | Injection port | Add (replace) test media |
1) Forward detection
Open the plugs of output 1 and output 2, open the stabilizing valve, adjust the fine adjustment to the appropriate position to lock the air release valve, gently operate the lever to bring the pressure guiding liquid to the output port, install the standard gauge and the calibrated gauge, operate the lever to increase the pressure to near the calibration point, use the fine adjustment to approach the calibration point, and detect point by point until full range.
2) Return inspection
According to the regulations, after two readings at full range, slowly reduce the pressure to the calibration point with fine adjustment, and check point by point until zero. When the fine adjustment cannot reduce the pressure, gently open the relief valve to reduce the pressure.
When the pressure drops slowly under high pressure, the stabilizing valve can be locked.
4. Precautions
1) The pressure pump should be placed on a flat and easy to operate table to avoid entering environments with highly corrosive gases and dust particles.
2) The tested instrument should be cleaned appropriately to prevent stains from entering the detection circuit.
3) Output 1 and output 2 are dedicated fast interfaces with sealed end faces. When installing the instrument, there is no need to use any tools. Cut evenly and tighten firmly. When the sealing surfaces are tightly attached to each other, tighten 2-3 more turns.
4) Regularly check the lower locking nut to ensure that the connection is secure. If the connection is not reliable, it may cause detachment when subjected to high pressure.
5) When the pressure generated by the lever is heavy, it is easier to use fine-tuning to increase the pressure.
Product Display
Online inquiry
