ZY-01Type online atmospheric pre concentration device
1、 Instrument Introduction
To meet the requirements of the online monitoring system for volatile organic compounds in the supporting environment airVOCThe detection limit of s has been reachedppbLevel, independently developed and launched by the companyZY-01Type online atmospheric pre concentration deviceIt can seamlessly integrate with online monitoring systems from various manufacturers.
2、 InstrumentCharacteristics and main functions
1. Color touch screen display, controlled by microcomputer program, with main functions including:
⑴ Method parameter setting, real-time display of working status, and running time;
⑵ The desorption zone, injection valve, sample transfer tube, and secondary desorption zone are all separately heated and temperature controlled;
⑶ Load the applicable analysis program and automatically complete the processing of the entire sample;
⑷Can be increased according to user needsDouble cold trapcomponent;
Equipped with dual interfaces: receiving sampling system output signal interface and sending start signal interface;
2. Adopting electronic refrigeration technology, convenient and practical;
3Through time programming, automatic functions such as adsorption, desorption, low-temperature secondary adsorption, desorption, injection, and blowback cleaning can be achieved;
4The sample transfer tube and injection valve treated with inertization have automatic blowback function, which avoids cross contamination of different samples;
3、 InstrumentMain technologiesparameter
1. desorption1Temperature control range:
Room temperature~400℃ In increments1℃Appointment;
2. Temperature control range of valve injection system:
Room temperature~240℃Increase by 1℃Appointment;
3. Temperature control range of sample transfer pipeline:
Room temperature~240℃Increase by 1℃Ren Shi
(For operational safety, the temperature control of the transmission pipeline adopts low-voltage power supply);
4. desorption2Temperature control range:
150~400℃ In increments1℃Appointment;
(Heating rate)2000℃/min)
5. Temperature control range of cold trap:
-35℃~50℃In increments1℃Ren Shi
(Using the most advanced electronic refrigeration device, no need for liquid nitrogen refrigeration)
6. Temperature control accuracy:< ±0.1℃;
7. Desorption recovery rate:>98%(Related to components);
8. Sample desorption time:0.01min~500min;
9. Sample blowing time:0.01min~500min;
10Sample injection time: 0.01min~500min;
11Backblowing time of pipeline: 0.01min~500min;
12.Backblowing cleaningTraffic:0~100ml/Min (continuously adjustable);
13Effective size of the instrument: 500×482×222mm3;
14Weight of the instrument: approximately20kg。